Technical Manual
5-8 Welch-Allyn, Inc.
70898-0000A
Field Serviceable Repairs
Repairs are considered field serviceable if the repair will not alter the calibration or
proper operation of the instrument.
All probes designed to work with the thermometer are fully interchangeable.
All components in the Model 678 and Model 679 can be replaced without affecting
instrument operation or calibration. Some minor changes to the exact calibration
point will be caused by changing R8, and R13, but as long as the proper type and
tolerance resistors are used (0.1% and 0.05% as supplied by Welch Allyn), the
unit will remain within specifications.
Replacement of the LCD frame is somewhat difficult due to the need to assemble
it while under pressure to assure proper compression of the elastomeric
connector.
Note:
Do not glue the LCD frame to the display PCB if the frame latching pins are
broken. This will destroy the display. Replace with a new LCD frame.
Field Serviceable Parts
All parts are serviceable by qualified technicians.
Note:
The following parts lists are current as of the date of publication. Parts and part
numbers may be changed without notice. Check with Welch-Allyn prior to
ordering parts to verify current parts for any replacement requirement.