Command and Response Strings: Examples SureTemp Plus Module OEM Implementation
60 Confidential Welch Allyn OEM Technologies
(3503) Lock Thermometer Probe
Locks or unlocks access to the module’s Thermometer probe switch. Once locked, the probe
switch may be turned on or off via the “Set Temperature Probe” command.
Example:
Command String 03 02 10 35 03 01 01 00 01 B0
Preamble 03 02 10
Command Code 35 03
Type code 01 (command)
Data size 01 00 (1 byte being sent)
Data 01 (where 1 locks probe switch, 0 unlocks probe switch)
Checksum B0
Response String 03 02 10 35 03 02 00 00 B1
Preamble 03 02 10
Command Code 35 03
Type Code 02 (good response)
Data Size 00 00 (0 bytes returned)
Data none
Checksum B1