SureTemp Plus Module OEM Implementation Communications Protocol
Welch Allyn OEM Technologies Confidential 43
(2A35) Set Thermometer Probe Mode
Causes the module to set the current probe mode from the host.
Command Code (hex)
2A35
Data
Size (bytes) Type Description
1 int8 0: Oral
1: Pediatric Axillary
2: Adult Axillary
Response Data
Size (bytes) Type Description
0 N/A None
Notes
The probe mode controls whether the oral, pediatric axillary, or adult axillary algorithm is to
be used with an oral/axillary probe. This command has no effect if the current probe type is
rectal.