Welch-allyn Connectivity Server - Programmers guide Manual do Utilizador Página 13

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 64
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 12
Programmer’s Guide System overview 9
Data exported from the Acuity System to WACS
The Acuity System exports the following vital-sign data:
Table 4. Numeric vital-sign data exported from Acuity Systems to WACS
Vital sign Units Default units Preferred HL7 tag
(OBX-3)
Heart rate BPM (1/min) BPM (1/min) Heart Rate
Temperature 1, 2 DegF (°F),
DegC (°C)
DegC (°C) Temperature
SpO
2
Per cent (%) Per cent (%) SPO2
CO
2
In mmHg, kPa, % mmHg CO2(In)
CO
2
Ex mmHg, kPa, % mmHg CO2(Ex)
RESP Br/M (1/min) Br/M (1/min) RR/BR
NIBP (DIA / SYS / MEAN) mmHg mmHg NIBP
IBP - 1 (DIA / SYS / MEAN)
unlabeled
arterial
intracranial
umbilical artery
umbilical vein
pulmonary artery
central venous
mmHg mmHg
IBP1
ART
ICP
UA
UV
PA
CVP
IBP - 2 (DIA / SYS / MEAN)
unlabeled
arterial
intracranial
umbilical artery
umbilical vein
pulmonary artery
central venous
mmHg mmHg
IBP2
ART
ICP
UA
UV
PA
CVP
PVCRate PVC/Min (1/min) PVC/Min (1/min) PVC