Welch-allyn Acuity and Mobile Acuity LT Central Monitoring Sys Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Equipamento Welch-allyn Acuity and Mobile Acuity LT Central Monitoring Sys. Welch Allyn Acuity and Mobile Acuity LT Central Monitoring Systems - User Manual User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Central Monitoring Systems

Acuity®andMobile Acuity LTCentral Monitoring SystemsDirections for useSoftware version 8.2x

Página 2

6 Chapter 1 General information Welch Allyn Acuity Central Monitoring SystemCautionsWARNING Whenever an interruption in Acuity System CPU operation oc

Página 3

96 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring System

Página 4

697Review or print patient dataReview patient dataIf your Acuity System includes the Full Disclosure option, waveforms and trendinformation for all pa

Página 5 - General information

98 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemTo view the patient’s data1. In the Patient List window or Other

Página 6 - Conventions

Directions for use Chapter 6 Review or print patient data 99Review graphical trendsIn the Graphical Trends Review window, you can review and print dat

Página 7 - General warnings and cautions

100 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemAdjust the graphical trends display and locate eventsTo adjust

Página 8

Directions for use Chapter 6 Review or print patient data 101Review tabular trends and eventsIn the Tabular Trends and Events Review window, you can r

Página 9

102 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemAdjust the content of the tabular listTo adjust the time period

Página 10 - Cautions

Directions for use Chapter 6 Review or print patient data 103Adjust the waveform displayTo display waveform(s) that correspond to a trend or event in

Página 11

104 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemReview waveforms(Full Disclosure option)In the Waveforms Review

Página 12 - Related documents

Directions for use Chapter 6 Review or print patient data 105Display waveforms and eventsTo select and adjust waveforms1. Check up to five boxes to di

Página 13 - Acuity System overview

Directions for use Chapter 1 General information 7Acuity System compliance with safety standardsThe CE Mark and Notified Body Registration Number sign

Página 14 - Diagram of the System

106 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemTo locate and display alarm and event information1. Along the t

Página 15

Directions for use Chapter 6 Review or print patient data 107Review arrhythmia events(Arrhythmia Analysis option)In the Arrhythmia Events Review windo

Página 16

108 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemOpen an Arrhythmia Events Review windowTo open an Arrhythmia Ev

Página 17

Directions for use Chapter 6 Review or print patient data 109Adjust the content of both list sectionsTo adjust the time period covered in both list se

Página 18

110 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemAdjust the waveform displayTo display waveform(s) that correspo

Página 19 - Acuity System HA CPU

Directions for use Chapter 6 Review or print patient data 111Delete or undelete an arrhythmia event marker from review windows and printouts(Advanced

Página 20 - Central station screen

112 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemReview ST analysis informationThis section explains how to revi

Página 21

Directions for use Chapter 6 Review or print patient data 113Review OxyCRG information(Full Disclosure option)In the OxyCRG review window, you can rev

Página 22 - System screen views

114 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemAdjust the OxyCRG display and locate alarms.To adjust the time

Página 23 - System buttons

Directions for use Chapter 6 Review or print patient data 115Print and review printoutsThe Acuity System includes six types of printouts:• Snapshot pr

Página 24

8 Chapter 1 General information Welch Allyn Acuity Central Monitoring SystemRelated documentsAcuity and Mobile Acuity LT Central Monitoring SystemsDir

Página 25

116 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemPrint a Snapshot printout on demandTo print a snapshot from a W

Página 26 - Waveform Window

Directions for use Chapter 6 Review or print patient data 117Review printoutsReview printouts capture information shown in a current or non-monitored

Página 27 - Virtual Monitor

118 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemRead a review printoutTo read a review printoutA. Type of print

Página 28

Directions for use Chapter 6 Review or print patient data 119Alarm printoutsIf Print on Alarm is set for a patient’s parameter and the parameter alarm

Página 29 - Admit and begin monitoring

120 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemReview an alarm printoutA. Patient name appears if patient is c

Página 30

Directions for use Chapter 6 Review or print patient data 121Patient List printoutsYou can print lists of patients in your unit. If your system is par

Página 31 - Readmit a discharged patient

122 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring System12-lead printoutsYou can perform 12-lead ECG interpretation and

Página 32

Directions for use Chapter 6 Review or print patient data 123Review a 12-lead reportA. Patient information appears if patient is confirmed in the Acui

Página 33

124 Chapter 6 Review or print patient data Welch Allyn Acuity Central Monitoring SystemHemodynamic Report printoutsHemodynamic Reports are a collectio

Página 34

7125Move a monitored patientYou can move a monitored patient in these ways:• Move a patient and monitor to a new room within the unit or transfer to a

Página 35 - Micropaq monitor

29Acuity System overviewSystem configurationsThe Acuity System consists of an Acuity Central Station that receives, displays andanalyzes information f

Página 36

126 Chapter 7 Move a monitored patient Welch Allyn Acuity Central Monitoring SystemArrange a patient move from the Acuity Central StationIf you want t

Página 37 - Propaq LT monitor

Directions for use Chapter 7 Move a monitored patient 127Move patients and their hardwired monitorsTo disconnect, move and reconnect a hardwired monit

Página 38

128 Chapter 7 Move a monitored patient Welch Allyn Acuity Central Monitoring SystemTo switch a hardwired Propaq CS monitor to a wireless connectionThr

Página 39 - Wireless Propaq CS monitor

Directions for use Chapter 7 Move a monitored patient 129Move patients and their wireless monitorsArrange a patient move from the patient monitorTo ar

Página 40

130 Chapter 7 Move a monitored patient Welch Allyn Acuity Central Monitoring SystemTo arrange a patient move from a wireless Propaq CS monitor1. From

Página 41

Directions for use Chapter 7 Move a monitored patient 131Move the patient to a different monitorYou can move patients who are connected to hardwired o

Página 42

132 Chapter 7 Move a monitored patient Welch Allyn Acuity Central Monitoring System5. Enter or confirm the patient information.For instructions, see “

Página 43

8133Discontinue Acuity SystemmonitoringStop monitoring a patientYou can either temporarily interrupt or permanently stop patient monitoring.For inform

Página 44 - Patient ID window 1

134 Chapter 8 Discontinue Acuity System monitoring Welch Allyn Acuity Central Monitoring SystemWARNING If your system includes arrhythmia analysis, an

Página 45 - Patient ID window 2

Directions for use Chapter 8 Discontinue Acuity System monitoring 135Permanently stop monitoringTo permanently stop monitoring a patient1. From the mo

Página 46

10 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemDiagram of the SystemA. Hallway message panel Optional panels, located

Página 47 - 2 spot check

136 Chapter 8 Discontinue Acuity System monitoring Welch Allyn Acuity Central Monitoring SystemFor instructions, see “Discharge a patient from Acuity

Página 48

9137Status iconsStatus icon colorsThe color of a status icon conveys the status of the patient, patient monitor or CentralStation component that the i

Página 49

138 Chapter 9 Status icons Welch Allyn Acuity Central Monitoring SystemPatient/monitor status iconsPatient/Monitor status icons show status and act as

Página 50

Directions for use Chapter 9 Status icons 139Patient/monitor status icons and recommended actionThis table illustrates these things about patient/moni

Página 51

140 Chapter 9 Status icons Welch Allyn Acuity Central Monitoring SystemAlarm Tone Suspendedwith No ViolationBeneath alertsand alarmsLocations represen

Página 52

Directions for use Chapter 9 Status icons 141Too Many HardwiredMonitorsTakes priorityover low-levelalerts.Location representing theextra monitor.[room

Página 53

142 Chapter 9 Status icons Welch Allyn Acuity Central Monitoring SystemMonitorCommunication LostTake priorityover low-levelalerts.Locations representi

Página 54

Directions for use Chapter 9 Status icons 143Patient AlarmorTelemetry DropoutTake priorityover alerts.Locations representingmonitors that are audiblya

Página 55

144 Chapter 9 Status icons Welch Allyn Acuity Central Monitoring SystemAlarm Tone SuspendedOnly lethaleventsinterruptthe suspensionperiod. Allother ne

Página 56 - Arrhythmia Analysis option

Directions for use Chapter 9 Status icons 145System component status iconsAcuity System component icons do these things:• Show component status• Open

Página 57

Directions for use Chapter 2 Acuity System overview 11Networked Acuity System featuresNetworked Acuity Systems include various features. This section

Página 58 - Arrhythmia event definitions

146 Chapter 9 Status icons Welch Allyn Acuity Central Monitoring SystemSystem component status icons and recommended actionThis table illustrates thes

Página 59

Directions for use Chapter 9 Status icons 147Central PrinterProblemCentral Printers OKLow-level alert Yellow icon: appears atAcuity Central Stationare

Página 60

148 Chapter 9 Status icons Welch Allyn Acuity Central Monitoring SystemNo Devices RunningClinician Notifier areReachableLow-level alert Acuity Central

Página 61

10149Troubleshooting and maintenanceContact informationIf you are unable to find the answer to a question, or if you encounter a problem thatcannot be

Página 62

150 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemContentsTroubleshootingCentral station displays ...

Página 63 - ♥. However

Directions for use Chapter 10 Troubleshooting and maintenance 151TroubleshootingCentral station displaysWARNING Replace central station hardware withi

Página 64

152 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemConnection, patient room and patient IDNo audible alarm or

Página 65 - ♥ appears in the

Directions for use Chapter 10 Troubleshooting and maintenance 153Sometimes patients are assigned an IDnumber that I didn’t give them.If you do not ent

Página 66

154 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemEquipment alertsTable 12. Troubleshooting equipment alertsP

Página 67

Directions for use Chapter 10 Troubleshooting and maintenance 155Alarm limitsA Cuff Not Detected or a No PulsesDetected equipment alert is occurring a

Página 68

12 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemTo open a Waveform Window on a PLR station screenPLR stations are typi

Página 69

156 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemPatient alarm indicationsTable 14. Troubleshooting patient

Página 70

Directions for use Chapter 10 Troubleshooting and maintenance 157A false alarm occurred for Heart rate,ECG, ST analysis or arrhythmia analysis.WARNING

Página 71

158 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemAudible alarm or alert tones are delayedat the patient moni

Página 72

Directions for use Chapter 10 Troubleshooting and maintenance 159Heart rate and ECG displaysTable 15. Troubleshooting heart rate and ECG displaysProbl

Página 73 - ST Analysis option

160 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemCO2, SpO2, IPI, and NIBP displaysTable 16. Troubleshooting

Página 74 - E. Reference beat (green)

Directions for use Chapter 10 Troubleshooting and maintenance 161Arrhythmia Analysis and ST AnalysisTable 17. Troubleshooting arrhythmia and ST-segmen

Página 75 - Turn on or off ST analysis

162 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemProblem ExplanationWARNING In Acuity Systems that include a

Página 76

Directions for use Chapter 10 Troubleshooting and maintenance 163PrintingTable 18. Troubleshooting printingProblem ExplanationThere is a delay in the

Página 77 - J-points (green)

164 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemMaintenanceAdjust Acuity System timeIf your facility had an

Página 78 - and Print on

Directions for use Chapter 10 Troubleshooting and maintenance 165Identify or change a monitor’s network name (ESSID)This section describes how to iden

Página 79

Directions for use Chapter 2 Acuity System overview 13Flexible patient access (FPA) using virtual network computing (VNC)The flexible patient access o

Página 80

166 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemIdentify or change a Micropaq monitor network nameTo identi

Página 81

Directions for use Chapter 10 Troubleshooting and maintenance 167Identify or change a Propaq LT monitor network nameTo identify a Propaq LT monitor ne

Página 82

168 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemIdentify or change a Wireless Propaq CS network nameTo iden

Página 83 - Edit patient information

Directions for use Chapter 10 Troubleshooting and maintenance 169Mobile Acuity LT Systems: Adjust localization parametersIf your Acuity System is a mo

Página 84

170 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemPower outagesTo turn off the Acuity System1. Notify staff m

Página 85 - Patient Note

Directions for use Chapter 10 Troubleshooting and maintenance 171To turn on the Acuity System1. Turn on the power to these components, in this sequenc

Página 86

172 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring SystemPeriodically check equipment for safetyThe Acuity System ne

Página 87 - Respond to alarms and alerts

Directions for use Chapter 10 Troubleshooting and maintenance 173Equipment disposalAcuity Central Monitoring SystemDo not dispose of this product as u

Página 88 - Suspend Area

174 Chapter 10 Troubleshooting and maintenance Welch Allyn Acuity Central Monitoring System

Página 89 - Related to indications

11175SpecificationsMonitor/Acuity System compatibilityGeneral system monitoring capabilitiesTable 19. Patient monitor/Acuity System compatibilityPatie

Página 90 - Related to NIBP alarms

14 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring System4. Enter your VNC client password.If you are using the VNC Enterprise

Página 91 - The latched alarm

176 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemTable 21. Networked Acuity System capabilitiesFunction Primary AcuitySystemA

Página 92 - Respond to a DROPOUT alert

Directions for use Chapter 11 Specifications 177Acuity System heart rateTable 22. Acuity System heart rate calculationAcuity System configuration Meth

Página 93

178 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemAcuity System pacemaker detectionTable 24. Pacemaker detection with the Acui

Página 94

Directions for use Chapter 11 Specifications 179Acuity System alarm and alert indicationsTable 25. General Acuity System alarm and alert indicationsFe

Página 95 - Respond to a RELEARN alert

180 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemMessage window(At top of Acuity CentralStation screen)Specific message match

Página 96

Directions for use Chapter 11 Specifications 181Table 27. Alarm and alert indications at the Acuity System Virtual MonitorAlarms(red)Monitor equipment

Página 97

182 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemTable 28. Alarm and alert indications at the Waveform WindowAlarms(red)Monit

Página 98

Directions for use Chapter 11 Specifications 183Table 29. Alarm and alert indications at (optional) hallway message panels and in messagewindow at the

Página 99 - if alarm level set to

184 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemElectromagnetic standards complianceThe Acuity Central Monitoring System com

Página 100

Directions for use Chapter 11 Specifications 185Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Acuity System is intended for use

Página 101 - Review or print patient data

Directions for use Chapter 2 Acuity System overview 15Paired system CPUs (High-availability pairs) and switchoversAcuity Systems that typically suppor

Página 102

186 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemGuidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Acuity S

Página 103 - Review graphical trends

Directions for use Chapter 11 Specifications 187Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and theAcuity

Página 104 - ♥ Arrhythmia event marker

188 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemAcuity System customizable defaults and factory defaultsAcuity System defaul

Página 105

Directions for use Chapter 11 Specifications 189Virtual Monitor and Waveform WindowTable 32. Virtual MonitorFeatures Customizable default Factory defa

Página 106 - Time line

190 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemSetup windowsHeart rate (HR) For patients being monitored for arrhythmias,th

Página 107 - •) and ventricular (v)

Directions for use Chapter 11 Specifications 191Table 35. Setup window—Arrhythmia AlarmsFeatures Customizable default Factory defaultAlarm level Mediu

Página 108 - Review waveforms

192 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemClicking Arr/ST Relearn or RELEARNClicking Analyze PacersClicking Single ECG

Página 109 - Display waveforms and events

Directions for use Chapter 11 Specifications 193Table 36. Setup window—ST AnalysisFeatures Customizable default Factory defaultST Analysis OnST analys

Página 110 - Cursor bar

194 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemPresets values Alarm limits: 1.0 mmAlarm level: LowMeasurement Offset: Auto

Página 111 - Review arrhythmia events

Directions for use Chapter 11 Specifications 195Review windowsTable 39. Review window—Graphical TrendsFeatures Customizable default Factory defaultFre

Página 112

ii Welch Allyn Acuity Central Monitoring System© 2013 Welch Allyn. All rights are reserved. To support the intended use of the product described in th

Página 113

16 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemCentral station screenDepending upon the selected screen view, the Acu

Página 114

196 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemTable 40. Review window—Tabular Trends/EventsFeaturesFactory defaultFrequenc

Página 115

Directions for use Chapter 11 Specifications 197Table 41. Review window—WaveformsFeatures Factory defaultDisplay priority of waveforms Up to five (ten

Página 116

198 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring SystemTable 42. Review window—Arrhythmia EventsFeatures Factory defaultWaveform di

Página 117 - Review OxyCRG information

Directions for use Chapter 11 Specifications 199Table 43. Review window—OxyCRGFeatures Factory defaultHR/PR and SpO2trend data Graphed as lines connec

Página 118

200 Chapter 11 Specifications Welch Allyn Acuity Central Monitoring System

Página 119 - Print and review printouts

201WarrantyLIMITED WARRANTY: ACUITY CENTRAL MONITORING SYSTEM AND ACCESSORIESThis product is sold by Welch Allyn under the warranties set forth in the

Página 120 - Print Area

202 Warranty Welch Allyn Acuity Central Monitoring System

Página 121 - Review printouts

203IndexAAccess point, ethernet 10Acuity Systemadjust time 164Central Station 10Central Station screen 16compatibility with monitors 175components 10c

Página 122 - Read a review printout

204 Index Welch Allyn Acuity Central Monitoring Systemfalse alarms 159Full or Lethal Only options 54learning period during 64, 78, 91, 162, 171lethal,

Página 123 - Alarm printouts

Directions for use Index 205check periodically 172ESSID, identifying or changing a 165Ethernet, access point 10Event definitions, Arrhythmia 54Event m

Página 124 - Review an alarm printout

Directions for use Chapter 2 Acuity System overview 17F. Screen view buttons Change the overall Acuity System screen view. They vary in appearanceacco

Página 125 - Patient List printouts

206 Index Welch Allyn Acuity Central Monitoring Systemincorrect 37Monitorclearing trends and alarm limits 28, 35compatibility with Acuity System 175co

Página 126 - 12-lead printouts

Directions for use Index 207outages 170supply, uninterruptible 3Preadmit patient 25Presets button 46, 60, 74Print Next button 50Print On Alarmarrhythm

Página 127 - Review a 12-lead report

208 Index Welch Allyn Acuity Central Monitoring SystemSpecifications 175alarm and alert 179EMC 184factory defaults 188general monitoring 175heart rate

Página 128 - Hemodynamic Report printouts

Directions for use Index 209WWACS 1, 10Warnings and cautionsarrhythmia analysis 66connecting monitor to Acuity System 37ST Analysis 69suspending alarm

Página 129 - Move a monitored patient

210 Index Welch Allyn Acuity Central Monitoring System

Página 130

18 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemSystem screen viewsYou can change the screen view of the Acuity Centra

Página 131

Directions for use Chapter 2 Acuity System overview 19System buttonsThe system buttons in the lower left corner of the Acuity Central Station screen p

Página 132

20 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemCentral station map and status iconsYour system includes a map that is

Página 133

Directions for use Chapter 2 Acuity System overview 21D. Patient/monitor status icon Icons appear at locations with active monitors. Click a map icon

Página 134

22 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemWaveform WindowA Waveform Window is an abbreviated Virtual Monitor.To

Página 135

Directions for use Chapter 2 Acuity System overview 23Virtual MonitorAn Acuity System Virtual Monitor represents a patient monitor and the information

Página 136

24 Chapter 2 Acuity System overview Welch Allyn Acuity Central Monitoring SystemTo open a patient’s Virtual MonitorTake one of these steps at the Acui

Página 137 - Discontinue Acuity System

325Admit and begin monitoringPreadmit a patient to Acuity System monitoringIf you know a patient’s ID number before the patient is admitted, you can s

Página 138

iiiContents1 - General information ...1Indications for use ...1General

Página 139 - Permanently stop monitoring

26 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemTo assign a room number for a preadmitted patient, via the Patient

Página 140

Directions for use Chapter 3 Admit and begin monitoring 27Readmit a discharged patientYou can readmit a patient to the Acuity System if the patient’s

Página 141 - Status icons

28 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemBegin patient monitor communication with the systemFor information

Página 142 - Patient/monitor status icons

Directions for use Chapter 3 Admit and begin monitoring 293. Attach the cables to the monitor last, to optimize waveform interpretation during theinit

Página 143

30 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemAt the central station, the patient map icon remains in the yellow

Página 144

Directions for use Chapter 3 Admit and begin monitoring 31Start communication between a wireless monitor and the systemFor information about patient m

Página 145

32 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemTo enter patient information1. If prompted, press to highlight the

Página 146

Directions for use Chapter 3 Admit and begin monitoring 33Propaq LT monitorTo start Propaq LT monitor communication with the Acuity System and connect

Página 147

34 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemTo enter patient information1. Specify whether existing patient da

Página 148

Directions for use Chapter 3 Admit and begin monitoring 35Wireless Propaq CS monitorTo start wireless Propaq CS monitor communication with the system1

Página 149 - System component status icons

iv Contents Welch Allyn Acuity Central Monitoring System7 - Move a monitored patient ...125Move the patient and monitor to

Página 150

36 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemTo enter patient information1. If you see the CONFIRM PATIENT? pro

Página 151

Directions for use Chapter 3 Admit and begin monitoring 37Important considerations: connecting a monitor to the Acuity SystemRelated to specific patie

Página 152

38 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemRelated to leads, cables and the ECG displayWARNING Ensure that th

Página 153 - Contact information

Directions for use Chapter 3 Admit and begin monitoring 39WARNING Ensure these things to minimize false HR, Arrhythmia Analysis andST Analysis alarms,

Página 154 - Contents

40 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring SystemEnter and confirm a patient’s ID or room at the centralstationA pa

Página 155 - Troubleshooting

Directions for use Chapter 3 Admit and begin monitoring 417. Click Next Page to enter additional patient demographics such as date of birth,gender, he

Página 156

42 Chapter 3 Admit and begin monitoring Welch Allyn Acuity Central Monitoring System

Página 157

443Adjust monitoring settings andpatient informationSet up typical monitoring at the Virtual MonitorThis table presents typical monitoring functions t

Página 158 - Equipment alerts

44 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemResponse time: SpO2and CO2Task Clicks at V

Página 159 - Alarm limits

Directions for use Chapter 4 Adjust monitoring settings and patient information 45Lead selection: ECG and respirationTask Clicks at Virtual Monitor No

Página 160 - Patient alarm indications

11General informationIndications for useThe Acuity®Central Monitoring Station is intended to be used by clinicians for the centralmonitoring of neonat

Página 161

46 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemAdjust a patient’s alarm limitsIn the Alar

Página 162

Directions for use Chapter 4 Adjust monitoring settings and patient information 47Important considerations: adjusting alarm limits\WARNING Patients fo

Página 163 - Heart rate and ECG displays

48 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemAdjust general printing settingsThis secti

Página 164 - , IPI, and NIBP displays

Directions for use Chapter 4 Adjust monitoring settings and patient information 49Adjust preferences for a patient’s central station printoutsThis sec

Página 165

50 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemCoordinate system print jobs and central s

Página 166

Directions for use Chapter 4 Adjust monitoring settings and patient information 51Adjust a patient’s waveform displaysYou can select waveforms for dis

Página 167 - Printing

52 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemArrhythmia Analysis optionThe Acuity Syste

Página 168 - Maintenance

Directions for use Chapter 4 Adjust monitoring settings and patient information 53relearn. An Arrhythmia Analysis function that causes the Acuity syst

Página 169

54 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemArrhythmia event definitionsThe Acuity Sys

Página 170

Directions for use Chapter 4 Adjust monitoring settings and patient information 55Lethal Arrhythmia Analysis option: events detectedTable 4. Definitio

Página 171

2 Chapter 1 General information Welch Allyn Acuity Central Monitoring SystemAcuity System software further processes data acquired from patient monito

Página 172

56 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemFull Arrhythmia Analysis option: additiona

Página 173

Directions for use Chapter 4 Adjust monitoring settings and patient information 57BradycardiaA HR less than or equal to the patient’sbradycardia alarm

Página 174 - Power outages

58 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemOpen an Arrhythmia Alarms Setup windowIn t

Página 175

Directions for use Chapter 4 Adjust monitoring settings and patient information 59Turn off alarming for all arrhythmia types except lethal arrhythmias

Página 176 - System components

60 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemAdjust a patient’s arrhythmia alarm limits

Página 177 - Equipment disposal

Directions for use Chapter 4 Adjust monitoring settings and patient information 61Adjust preferences for a patient’s arrhythmia alarmingTo adjust pref

Página 178

62 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring System• Print on Alarm: To specify that central

Página 179 - Specifications

Directions for use Chapter 4 Adjust monitoring settings and patient information 63Initiate relearn of a patient’s ECG rhythmDescription of the relearn

Página 180

64 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring System.WARNING In Acuity Systems that include an

Página 181 - Acuity System heart rate

Directions for use Chapter 4 Adjust monitoring settings and patient information 65To initiate relearn of a patient’s ECG rhythm for arrhythmia and ST-

Página 182

Directions for use Chapter 1 General information 3General warnings and cautionsWarningsRelated to system components and configurationWARNING Use only

Página 183

66 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemImportant considerations: Arrhythmia Analy

Página 184

Directions for use Chapter 4 Adjust monitoring settings and patient information 67Related to patient monitor heart rate indicationsRelated to electrod

Página 185

68 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemWARNING Carefully monitor lead-wire connec

Página 186

Directions for use Chapter 4 Adjust monitoring settings and patient information 69ST Analysis optionThe Acuity System ST Analysis option analyzes the

Página 187

70 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemExamine the ST Analysis Setup windowThe ST

Página 188

Directions for use Chapter 4 Adjust monitoring settings and patient information 71Turn on or off ST analysisTo turn on or off ST analysis1. In the pat

Página 189

72 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemView current beats and ST-segment deviatio

Página 190

Directions for use Chapter 4 Adjust monitoring settings and patient information 73Set analysis preferences for a patient’s ST analysisTo set analysis

Página 191

74 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemSet alarm preferences for ST AnalysisTo se

Página 192 - Central Station windows

Directions for use Chapter 4 Adjust monitoring settings and patient information 75Initiate a relearn of a patient’s reference beats for ST analysisDes

Página 193 - Table 32. Virtual Monitor

4 Chapter 1 General information Welch Allyn Acuity Central Monitoring SystemRelated to close physical surveillance and specific patient populationsWAR

Página 194 - Setup windows

76 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemTo initiate a relearn of a patient’s refer

Página 195

Directions for use Chapter 4 Adjust monitoring settings and patient information 77Important considerations: ST AnalysisWARNING Before you set customiz

Página 196 - (ST relearn)

78 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemWARNING In Acuity Systems that include an

Página 197 - based on

Directions for use Chapter 4 Adjust monitoring settings and patient information 79Edit patient informationYou can revise information within the Acuity

Página 198

80 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring SystemEdit a patient ID number: add an amended (

Página 199 - Review windows

Directions for use Chapter 4 Adjust monitoring settings and patient information 81Add or remove a note in a patient’s Waveform WindowIf you would like

Página 200

82 Chapter 4 Adjust monitoring settings and patient information Welch Allyn Acuity Central Monitoring System

Página 201

583Respond to alarms and alertsOverview of central station alarm and alert indications• Alarms indicate patient vital-sign violations by showing red w

Página 202

84 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring SystemSuspend an audible alarm or alert tone for 90 secondsAll patient

Página 203

Directions for use Chapter 5 Respond to alarms and alerts 85Important considerations: suspending an audible alarm or alert toneRelated to indicationsW

Página 204

Directions for use Chapter 1 General information 5Related to monitoring interruptionWARNING Do not use the Arrhythmia Analysis or ST Analysis feature

Página 205 - Warranty

86 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring SystemRelated to heart rate (HR) and arrhythmia alarmsRelated to NIBP

Página 206

Directions for use Chapter 5 Respond to alarms and alerts 87Remove a latched arrhythmia alarmIf your Acuity System is configured to allow latched alar

Página 207

88 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring SystemRespond to specific alarms and alertsThis section describes a fe

Página 208

Directions for use Chapter 5 Respond to alarms and alerts 89Dismiss a CHECK PATIENT HR/PR alertA patient monitor uses a single lead to calculate heart

Página 209 - Directions for use Index 205

90 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring SystemDismiss an arrhythmia or ST alarm messageWhen you suspend an arr

Página 210

Directions for use Chapter 5 Respond to alarms and alerts 91Respond to a RELEARN alertWARNING In Acuity Systems that include an Arrhythmia option, a r

Página 211 - Directions for use Index 207

92 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring SystemTo respond to a RELEARN alert1. In the patient’s Virtual Monitor

Página 212

Directions for use Chapter 5 Respond to alarms and alerts 93Respond to a CHECK LEADS alertThe visual CHECK LEADS alert appears when the number of PVCs

Página 213 - Directions for use Index 209

94 Chapter 5 Respond to alarms and alerts Welch Allyn Acuity Central Monitoring SystemRespond to an arrhythmia NOISE alertThe visual NOISE alert occur

Página 214

Directions for use Chapter 5 Respond to alarms and alerts 95Summary of Acuity System alarm and alert levelsAlarm and alert levels determine which alar

Comentários a estes Manuais

Sem comentários