Propaq® CSVital Signs MonitorDirections for UseModels 242, 244, 246Software version 3.7X

Página 2

6 General information Welch Allyn Propaq CS Vital Signs MonitorIt is possible for the monitor to detect a problem that prevents the monitor from opera

Página 3 - Contents

96 Reference Welch Allyn Propaq CS Vital Signs MonitorQRS Detector Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)Neonatal Amplitude Range: 0

Página 4

Directions for Use Reference 97Real-Time ECG analog/defib sync specificationsSpecial cables are required to interface the defib sync connector to a Ph

Página 5 - General information

98 Reference Welch Allyn Propaq CS Vital Signs MonitorInvasive pressure specificationsCharacteristic SpecificationTransducer Type Strain-gauge resisti

Página 6

Directions for Use Reference 99NIBP specificationsCharacteristic SpecificationMethod OscillometricControl Automatic and manual measurement controlAuto

Página 7 - General warnings and cautions

100 Reference Welch Allyn Propaq CS Vital Signs MonitorTemperature specificationsCharacteristic SpecificationRange 0° to +50°C; 32° to +122°FDisplays

Página 8

Directions for Use Reference 101Pulse oximetry (SpO2) specificationsPulse Oximetry (SpO2) Specifications for Masimo SpO2Characteristic SpecificationSa

Página 9

102 Reference Welch Allyn Propaq CS Vital Signs MonitorPulse oximetry (SpO2) specifications for Nellcor SpO2Characteristic Specificationaa. When perfo

Página 10

Directions for Use Reference 103Capnography (CO2) specificationsGeneral CO2 Specifications (mainstream CO2 and sidestream CO2)Characteristic Specifica

Página 11 - Controls and connectors

104 Reference Welch Allyn Propaq CS Vital Signs Monitor Mainstream CO2 specificationsCharacteristic SpecificationMainstream CO2 sensorSensor Type Main

Página 12 - COMPATIBLE

Directions for Use Reference 105Sidestream CO2 specificationsCharacteristic SpecificationSensor Type Sidestream, internalPrinciple of Operation Non-di

Página 13 - Option connectors

Directions for Use General information 7Controls and connectorsTouch-screen controlsThe front panel touch-screen provides five softkeys along the bott

Página 14

106 Reference Welch Allyn Propaq CS Vital Signs MonitorAlarms specificationsNurse Call specificationsNurse Call cable specificationsThis cable (008-06

Página 15 - Main Menu

Directions for Use Reference 107Trends specificationsDisplay specificationsCharacteristic Specification aa. Assumes SpO2 and CO2 functions are present

Página 16 - Setup Menus

108 Reference Welch Allyn Propaq CS Vital Signs MonitorWireless Propaq CS Radio specifications (5 GHz)Channel restrictions in the 5-GHz band, by count

Página 17

Directions for Use Reference 109Wireless Propaq CS Radio specifications (2.4 GHz)Monitor (environmental) specificationsCharacteristic SpecificationFle

Página 18

110 Reference Welch Allyn Propaq CS Vital Signs MonitorMonitor (physical) specificationsCharacteristic SpecificationProtection classifications, all co

Página 19

Directions for Use Reference 111Printer specificationsCharacteristic SpecificationOperationOperating Modes Continuous, Snapshot, Auto Print, Auto Tren

Página 20 - ? ? ? is displayed

112 Reference Welch Allyn Propaq CS Vital Signs MonitorPower SpecificationsCharacteristic SpecificationMode of Operation ContinuousBattery Pack Type S

Página 21

Directions for Use Reference 113Power Adapter SpecificationsCharacteristic SpecificationProtection classifications, all adapters a ba. Per EN 60601-1

Página 22

114 Reference Welch Allyn Propaq CS Vital Signs Monitor

Página 23 - Directions for Use Setup 19

9115ComplianceGeneralThe 802.11a Wireless PC Card must be installed and used in strict accordance with the manufacturer’s instructions as described in

Página 24

8 General information Welch Allyn Propaq CS Vital Signs MonitorPatient connectors (left side panel)NIBPPSNIECG / EKG RESPINV. BPT1T2P1!NIBPINV. BPP1

Página 25 - Monitoring

116 Chapter 9 Compliance Welch Allyn Micropaq Monitorsubstitution or attachment of connecting cables and equipment other than specified by Welch All

Página 26 - Ω series resistors

Directions for Use Chapter 9 Compliance 117Electromagnetic compatibilitySpecial precautions concerning electromagnetic compatibility (EMC) must be t

Página 27

118 Chapter 9 Compliance Welch Allyn Micropaq MonitorPropaq CS Monitor: Guidance and manufacturer’s declaration—Electromagnetic immunityThe Propaq C

Página 28

Directions for Use Chapter 9 Compliance 119Propaq CS Monitor: Guidance and manufacturer’s declaration—Electromagnetic immunityThe Propaq CS monitor

Página 29 - EXT = Extended

120 Chapter 9 Compliance Welch Allyn Micropaq MonitorRecommended separation distances between portable and mobile RF communications equipment and th

Página 30

121IndexAAccess points, 72AC Power AdapterChange fuses, 83Change voltage selection, 83Connect to monitor, 81AC power adapter, specifications, 113AC po

Página 31 - Pacemaker signal indicators

122 Index Welch Allyn Propaq CS Vital Signs MonitorDate format setting, 92Decimal character setting, 92Default inflation pressure (NIBP), 32Default se

Página 32

Directions for Use Index 123Equipment alert messages, 57Monitoring, 28Numeric format adjustment, 31Range Mode, 30Rescale Mode, 30Specifications, 98Tra

Página 33

124 Index Welch Allyn Propaq CS Vital Signs MonitorSymbol in display, 34, 66Symbol in printout, 65TURBOCUF repeated measurements, 35Warnings, 32NIBP I

Página 34

Directions for Use Index 125Set time and date, 91Settings, factory default, 93SetupAlarm limits, 17Alarms, 17Custom patient modes, 19New patient, 15Pa

Página 35 - Second format

Directions for Use General information 9Option connectors!CO2!CO2Mainstream CO2 Connector Sidestream CO2 ConnectorNurse Call ConnectorNellcor SpO2 Con

Página 36

126 Index Welch Allyn Propaq CS Vital Signs MonitorUUnits of measure, CO2, 92Units of measure, temperature, 36, 92VVentilation, high frequency and RES

Página 37 - Take the NIBP reading

10 General information Welch Allyn Propaq CS Vital Signs MonitorDisplayYou can select up to four waveforms to be shown on the Propaq CS monitor. When

Página 38

Directions for Use General information 11MenusMenus for some patient vital signs are displayed only if the option is included in your Propaq CS monito

Página 39

12 General information Welch Allyn Propaq CS Vital Signs MonitorSetup Menus3Service menu tests are only for use by authorized service personnel and ar

Página 40

Directions for Use General information 13Learn Propaq CS operation with Inservice ModeYou can practice using the Propaq CS monitor with the in-service

Página 41 - monitoring

14 General information Welch Allyn Propaq CS Vital Signs MonitorNoteThe inservice mode is not available if the monitor detects that a sensor has been

Página 42

215SetupPrepare the Propaq CS monitor for a new patient1. Press the gray recessed MONITOR button on the right side of the monitor to turn the monitor

Página 43

ii Welch Allyn Propaq CS Vital Signs Monitor© 2008 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, t

Página 44

16 Setup Welch Allyn Propaq CS Vital Signs MonitorWhenever you change the patient mode, the alarm limit settings, maximum NIBP cuff inflation pressure

Página 45 - “spot-check” monitoring

Directions for Use Setup 17Set patient alarms and alarm limits1. From the Main Menu, press SETUP, ALARMS to access the Alarms Status Menu:2. Press LIM

Página 46 - Perform Mainstream CO

18 Setup Welch Allyn Propaq CS Vital Signs Monitor3. Press NEXT PARAMETER to highlight the parameter you want to change, then press NEXT SETTING to hi

Página 47

Directions for Use Setup 19Change Powerup Patient Mode or store customized settingsThe Propaq CS monitor has standard, preset, default powerup setting

Página 48 - Perform sidestream CO

20 Setup Welch Allyn Propaq CS Vital Signs MonitorCustomize Patient Mode settings1. From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the

Página 49

321MonitoringPerform ECG/RESP monitoringWARNING Impedance pneumography detects respiratory effort via changes in chest volume; therefore, impedance

Página 50 - display and alarm limits

22 Monitoring Welch Allyn Propaq CS Vital Signs Monitor• Impedance pneumography (RESP) is not recommended for use with high frequency ventilation.• Si

Página 51 - (not mainstream CO

Directions for Use Monitoring 23• Severe artifact and interference (such as defibrillation interference) can cause the waveform to move off the displa

Página 52

24 Monitoring Welch Allyn Propaq CS Vital Signs Monitor7. Plug the ECG cable into the ECG connector on the monitor's left side panel.8. Support t

Página 53 - Alarms & Alerts

Directions for Use Monitoring 2513. Press MORE to display the second ECG/RESP menu and status window:014. Press NEXT and CHANGE as desired to adjust t

Página 54 - 1. P r e s s Silence/Reset

iiiContents1 - General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Intended use. . . . . . . . . . . . .

Página 55 - 4 MIN SUSPND

26 Monitoring Welch Allyn Propaq CS Vital Signs Monitor15. Set alarms according to your facility’s standards.Use the ECG filter to display a better wa

Página 56

Directions for Use Monitoring 27Using the Propaq CS monitor with pacemaker patientsIf the patient being monitored has a pacemaker, the Propaq CS monit

Página 57 - Station messages

28 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform Invasive Blood Pressure (IBP) monitoring1. Inspect the transducer cable and transducer

Página 58

Directions for Use Monitoring 295. If the ZERO menu is not displayed, from the Main Menu press INVASIVE PRESSURE, and then ZERO P1 (or ZERO P2). The w

Página 59 - Connect nurse call option

30 Monitoring Welch Allyn Propaq CS Vital Signs Monitor10. To display all invasive pressure waveforms on one scale (when two IBP channels are active),

Página 60 - RESP messages

Directions for Use Monitoring 3113. To change the format of the IBP numerics, from the second IBP menu press FORMAT.To restore the first format, press

Página 61 - NIBP messages

32 Monitoring Welch Allyn Propaq CS Vital Signs MonitorTake a Noninvasive Blood-Pressure (NIBP) readingAt powerup, the Propaq CS monitor has an NIBP d

Página 62

Directions for Use Monitoring 33To enable the Smartcuf filter:• Connect the ECG leads to the patient and perform ECG monitoring during NIBP.• From the

Página 63 - Mainstream CO

34 Monitoring Welch Allyn Propaq CS Vital Signs Monitor3. Place the cuff on the limb.If possible, place it at the same level as the heart. If above th

Página 64 - Sidestream CO

Directions for Use Monitoring 356. If motion artifact such as shivering, coughing, or other motion interferes with NIBP readings, do the following:Pos

Página 65

iv Contents Welch Allyn Propaq CS Vital Signs Monitor7 - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 66 - Program alert message

36 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform temperature monitoring1. Place the temperature probe on the patient.2. Plug the probe c

Página 67 - Defibrillator alert message

Directions for Use Monitoring 37Perform SpO2 monitoringEach SpO2 sensor is designed for application to a specific site on the patient within a certain

Página 68

38 Monitoring Welch Allyn Propaq CS Vital Signs Monitorto the sensor, the patient is in cardiac arrest or shock, or the patient has hypotension, sever

Página 69 - Printing & trends

Directions for Use Monitoring 39Perform SpO2 monitoring with Nellcor option1. Attach the sensor to the patient according to the sensor manufacturer&ap

Página 70 - Display or print trends

40 Monitoring Welch Allyn Propaq CS Vital Signs Monitor9. If the C-LOCK function is desired, press C-LOCK to set it to ON.C-LOCK synchronizes the puls

Página 71 - Delete all patient trends

Directions for Use Monitoring 41Perform SpO2 “spot-check” monitoringThe SpO2 Standby Mode allows you to remove the SpO2 sensor from a patient without

Página 72

42 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform Mainstream CO2 monitoringThe capnography (CO2) option measures End-tidal CO2 (ETCO2), I

Página 73 - Print OxyCRG

Directions for Use Monitoring 433. Plug in the CO2 sensor cable to the mainstream CO2 connector on the monitor left side panel.4. See page 46 and set

Página 74

44 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform sidestream CO2 monitoringThe capnography (CO2) option measures End-tidal CO2 (ETCO2), I

Página 75

Directions for Use Monitoring 452. See page 46 and set up the CO2 display and alarm limits, and then continue this procedure with step 3.3. A. For a n

Página 76

11General informationIntended useThe Propaq CS monitor is intended to be used by skilled clinicians for multiparameter vital signs monitoring of neona

Página 77 - EKG x 1000

46 Monitoring Welch Allyn Propaq CS Vital Signs MonitorSidestream CO2 accuracy decreases if additional tubing is connected to the sample line. Avoid c

Página 78 - The Acuity Menu

Directions for Use Monitoring 472. Press RANGE to select the CO2 waveform scale or range.To ch a n g e C O2 units (mmHg, kPa, or %) see “Change the

Página 79 - Acuity connector

48 Monitoring Welch Allyn Propaq CS Vital Signs Monitor7. P r e s s SOURCE to change between mainstream CO2 and sidestream CO2 monitoring (if both o

Página 80 - Press to disconnect

449Alarms & AlertsDescription of alarm and alert tone patterns• Patient alarm tone pattern: one second on, two seconds off• Apnea alarm tone patte

Página 81 - Wireless connection to Acuity

50 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorRe-enable an alarm or alert tone before the 90-second silence period has elapsed1. P r

Página 82

Directions for Use Alarms & Alerts 51Inhibit alarm and alert tones for four minutes: 4 MIN SUSPNDAs you perform patient care, there might be occas

Página 83 - Directions for Use Acuity 79

52 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorInhibit alarm and alert tones indefinitely: ALL ALARMSYou can suspend potential or cur

Página 84

Directions for Use Alarms & Alerts 53Summary of alarm and alert keys and Acuity Central Station messagesThe following table summarizes Silence and

Página 85 - Maintenance

54 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorCustomize alarm limits based on patient’s current vital signsTo quickly set all alarm

Página 86 - Right Side Panel

Directions for Use Alarms & Alerts 55Alarm holdoffsTo help minimize false alarms, the monitor briefly delays or “holds off” triggering alarms for

Página 87 - Replace power adapter fuses

2 General information Welch Allyn Propaq CS Vital Signs MonitorSymbolsThe following symbols may appear on the Propaq CS monitor or accessories. They a

Página 88

56 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorTroubleshooting equipment alert messagesSimultaneous equipment alertsMULTIPLE. If mult

Página 89 - Install printer paper

Directions for Use Alarms & Alerts 57IBP messagesTRANSDUCER NOT DETECTED. The transducer connection is broken.TRANSDUCER SHORT CIRCUIT. This messa

Página 90

58 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitorchannel has not updated its calibration in 15 minutes, the channel will briefly deacti

Página 91 - Recycling monitor components

Directions for Use Alarms & Alerts 59Temperature messagesPROBE NOT DETECTED. This message occurs when the Propaq CS monitor has successfully measu

Página 92 - Extended storage precautions

60 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitorwaveform is displayed without range values. Replace the adapter or replace the sensor.

Página 93

Directions for Use Alarms & Alerts 61ALTIMETER NOT CALIBRATED - EEPROM. The sidestream CO2 option has not been calibrated. Refer the Propaq CS mon

Página 94

62 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorNetwork alert messages with AcuityCHECK ACUITY/SERIAL CONNECTION. The Propaq CS detect

Página 95 - Reference

Directions for Use Alarms & Alerts 63Printer alert messagesCHECK DOOR. The door on the bottom of the printer is open. Close the door to remove thi

Página 96

64 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitor

Página 97 - Factory default settings

565Printing & trendsPrint patient dataPrint the displayed waveformsPress SNAPSHOT or START/STOP. The Propaq CS monitor prints up to three of the d

Página 98

Directions for Use General information 3General warnings and cautionsFamiliarize yourself with all warnings and cautions before using the Propaq CS mo

Página 99 - Specifications

66 Printing & trends Welch Allyn Propaq CS Vital Signs MonitorDisplay or print trendsDisplay or print a single trend1. To display a patient data t

Página 100

Directions for Use Printing & trends 67Print multiple trends manually or automatically1. P r e s s SETUP, MORE, PRINTER from the Main Menu to di

Página 101

68 Printing & trends Welch Allyn Propaq CS Vital Signs MonitorSet printer options and automatic printing1. P r e s s SETUP, MORE, PRINTER from t

Página 102

Directions for Use Printing & trends 69Print OxyCRGThe OxyCRG is a printout of two minutes of continuous HR/PR and SpO2 numerics, and a compressed

Página 103 - NIBP specifications

70 Printing & trends Welch Allyn Propaq CS Vital Signs Monitor

Página 104 - Temperature specifications

671AcuityUnderstanding the Propaq CS, Acuity, and FlexNet networkThis section describes the operation of the Propaq CS monitor (software version 3.5X)

Página 105 - ) specifications

72 Acuity Welch Allyn Propaq CS Vital Signs MonitorThe Propaq CS monitor can communicate with an Acuity Central Station as part of Welch Allyn’s FlexN

Página 106

Directions for Use Acuity 73If the Wireless Propaq CS is moved out of range or loses communication with the FlexNet network and Acuity, it continues t

Página 107 - Capnography (CO

74 Acuity Welch Allyn Propaq CS Vital Signs MonitorThe Acuity MenuThe Acuity Menu on the Propaq CS allows you to control some of the interaction with

Página 108

Directions for Use Acuity 75Hardwired (Serial) connection to AcuityConnect the monitor to the Acuity System1. If the Propaq CS monitor has already bee

Página 109

4 General information Welch Allyn Propaq CS Vital Signs MonitorWARNING Do not connect more than one patient to a monitor. Do not connect more than o

Página 110 - Nurse Call specifications

76 Acuity Welch Allyn Propaq CS Vital Signs MonitorDisconnect the hardwired monitor from the Acuity System1. To disconnect the Propaq CS monitor from

Página 111 - Display specifications

Directions for Use Acuity 77Wireless connection to AcuityConnect a new patient1. Make sure the Acuity network cable is not plugged into the Acuity con

Página 112

78 Acuity Welch Allyn Propaq CS Vital Signs MonitorMonitor a patient out of range of AcuityWhile out of range of Acuity, the Wireless Propaq CS contin

Página 113

Directions for Use Acuity 79Reassign a monitored patient to a new room in the same unitIf a patient is being monitored and you want to assign them to

Página 114

80 Acuity Welch Allyn Propaq CS Vital Signs MonitorReassign the Wireless Propaq CS to a new patientIf you want to discontinue monitoring a patient and

Página 115 - Printer specifications

781MaintenanceAvoid electrostatic dischargeWhen humidity in the working environment decreases, the human body and other insulators can become charged

Página 116 - Power Specifications

82 Maintenance Welch Allyn Propaq CS Vital Signs Monitor2. Plug the ac adapter power cord into the ac power adapter and the ac power source outlet.3.

Página 117 - Power Adapter Specifications

Directions for Use Maintenance 83Replace power adapter fusesIf the green battery charging light is off and the ac power adapter does not provide power

Página 118

84 Maintenance Welch Allyn Propaq CS Vital Signs MonitorReplace monitor input power fuseIf the green battery charging light is off and the ac power ad

Página 119 - Compliance

Directions for Use Maintenance 85Install printer paper1. Lay the monitor on its back to gain access to the bottom of the printer.2. Squeeze the locks

Página 120 - European Union

Directions for Use General information 5WARNING If a product has been dropped or severely abused, send it to a qualified service person to confirm p

Página 121 - Electromagnetic compatibility

86 Maintenance Welch Allyn Propaq CS Vital Signs MonitorInspect and clean the monitor and accessoriesBefore cleaning, thoroughly inspect the monitor a

Página 122

Directions for Use Maintenance 87Service interval recommendationsAt the intervals recommended below, qualified biomedical service personnel should ser

Página 123

88 Maintenance Welch Allyn Propaq CS Vital Signs MonitorExtended storage precautionsRemoving the battery will erase all stored Custom patient mode set

Página 124

Directions for Use Maintenance 89The monitor displays a confirmation screen asking you to confirm that you want to change the network name.• If you pr

Página 125

90 Maintenance Welch Allyn Propaq CS Vital Signs Monitor

Página 126

891ReferenceSet the time and date1. From the Main Menu press SETUP, MORE, MORE to display the Time/Day window:2. Press NEXT, UP, and DOWN as needed to

Página 127 - Directions for Use Index 123

92 Reference Welch Allyn Propaq CS Vital Signs MonitorChange the date format, ECG filter, and units1. Make sure you are in the Adult patient mode (fro

Página 128

Directions for Use Reference 93Factory default settingsThe monitor is shipped from the factory with these preset default settings. For information abo

Página 129 - Directions for Use Index 125

94 Reference Welch Allyn Propaq CS Vital Signs MonitorHR Limits Adult: 50, 120 beats per minutePed: 50, 150 beats per minuteNeonate: 100, 200 beats pe

Página 130

Directions for Use Reference 95SpecificationsECG specificationsThe ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and A