Propaq® CSVital Signs MonitorDirections for UseModels 242, 244, 246Software version 3.7X
6 General information Welch Allyn Propaq CS Vital Signs MonitorIt is possible for the monitor to detect a problem that prevents the monitor from opera
96 Reference Welch Allyn Propaq CS Vital Signs MonitorQRS Detector Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)Neonatal Amplitude Range: 0
Directions for Use Reference 97Real-Time ECG analog/defib sync specificationsSpecial cables are required to interface the defib sync connector to a Ph
98 Reference Welch Allyn Propaq CS Vital Signs MonitorInvasive pressure specificationsCharacteristic SpecificationTransducer Type Strain-gauge resisti
Directions for Use Reference 99NIBP specificationsCharacteristic SpecificationMethod OscillometricControl Automatic and manual measurement controlAuto
100 Reference Welch Allyn Propaq CS Vital Signs MonitorTemperature specificationsCharacteristic SpecificationRange 0° to +50°C; 32° to +122°FDisplays
Directions for Use Reference 101Pulse oximetry (SpO2) specificationsPulse Oximetry (SpO2) Specifications for Masimo SpO2Characteristic SpecificationSa
102 Reference Welch Allyn Propaq CS Vital Signs MonitorPulse oximetry (SpO2) specifications for Nellcor SpO2Characteristic Specificationaa. When perfo
Directions for Use Reference 103Capnography (CO2) specificationsGeneral CO2 Specifications (mainstream CO2 and sidestream CO2)Characteristic Specifica
104 Reference Welch Allyn Propaq CS Vital Signs Monitor Mainstream CO2 specificationsCharacteristic SpecificationMainstream CO2 sensorSensor Type Main
Directions for Use Reference 105Sidestream CO2 specificationsCharacteristic SpecificationSensor Type Sidestream, internalPrinciple of Operation Non-di
Directions for Use General information 7Controls and connectorsTouch-screen controlsThe front panel touch-screen provides five softkeys along the bott
106 Reference Welch Allyn Propaq CS Vital Signs MonitorAlarms specificationsNurse Call specificationsNurse Call cable specificationsThis cable (008-06
Directions for Use Reference 107Trends specificationsDisplay specificationsCharacteristic Specification aa. Assumes SpO2 and CO2 functions are present
108 Reference Welch Allyn Propaq CS Vital Signs MonitorWireless Propaq CS Radio specifications (5 GHz)Channel restrictions in the 5-GHz band, by count
Directions for Use Reference 109Wireless Propaq CS Radio specifications (2.4 GHz)Monitor (environmental) specificationsCharacteristic SpecificationFle
110 Reference Welch Allyn Propaq CS Vital Signs MonitorMonitor (physical) specificationsCharacteristic SpecificationProtection classifications, all co
Directions for Use Reference 111Printer specificationsCharacteristic SpecificationOperationOperating Modes Continuous, Snapshot, Auto Print, Auto Tren
112 Reference Welch Allyn Propaq CS Vital Signs MonitorPower SpecificationsCharacteristic SpecificationMode of Operation ContinuousBattery Pack Type S
Directions for Use Reference 113Power Adapter SpecificationsCharacteristic SpecificationProtection classifications, all adapters a ba. Per EN 60601-1
114 Reference Welch Allyn Propaq CS Vital Signs Monitor
9115ComplianceGeneralThe 802.11a Wireless PC Card must be installed and used in strict accordance with the manufacturer’s instructions as described in
8 General information Welch Allyn Propaq CS Vital Signs MonitorPatient connectors (left side panel)NIBPPSNIECG / EKG RESPINV. BPT1T2P1!NIBPINV. BPP1
116 Chapter 9 Compliance Welch Allyn Micropaq Monitorsubstitution or attachment of connecting cables and equipment other than specified by Welch All
Directions for Use Chapter 9 Compliance 117Electromagnetic compatibilitySpecial precautions concerning electromagnetic compatibility (EMC) must be t
118 Chapter 9 Compliance Welch Allyn Micropaq MonitorPropaq CS Monitor: Guidance and manufacturer’s declaration—Electromagnetic immunityThe Propaq C
Directions for Use Chapter 9 Compliance 119Propaq CS Monitor: Guidance and manufacturer’s declaration—Electromagnetic immunityThe Propaq CS monitor
120 Chapter 9 Compliance Welch Allyn Micropaq MonitorRecommended separation distances between portable and mobile RF communications equipment and th
121IndexAAccess points, 72AC Power AdapterChange fuses, 83Change voltage selection, 83Connect to monitor, 81AC power adapter, specifications, 113AC po
122 Index Welch Allyn Propaq CS Vital Signs MonitorDate format setting, 92Decimal character setting, 92Default inflation pressure (NIBP), 32Default se
Directions for Use Index 123Equipment alert messages, 57Monitoring, 28Numeric format adjustment, 31Range Mode, 30Rescale Mode, 30Specifications, 98Tra
124 Index Welch Allyn Propaq CS Vital Signs MonitorSymbol in display, 34, 66Symbol in printout, 65TURBOCUF repeated measurements, 35Warnings, 32NIBP I
Directions for Use Index 125Set time and date, 91Settings, factory default, 93SetupAlarm limits, 17Alarms, 17Custom patient modes, 19New patient, 15Pa
Directions for Use General information 9Option connectors!CO2!CO2Mainstream CO2 Connector Sidestream CO2 ConnectorNurse Call ConnectorNellcor SpO2 Con
126 Index Welch Allyn Propaq CS Vital Signs MonitorUUnits of measure, CO2, 92Units of measure, temperature, 36, 92VVentilation, high frequency and RES
10 General information Welch Allyn Propaq CS Vital Signs MonitorDisplayYou can select up to four waveforms to be shown on the Propaq CS monitor. When
Directions for Use General information 11MenusMenus for some patient vital signs are displayed only if the option is included in your Propaq CS monito
12 General information Welch Allyn Propaq CS Vital Signs MonitorSetup Menus3Service menu tests are only for use by authorized service personnel and ar
Directions for Use General information 13Learn Propaq CS operation with Inservice ModeYou can practice using the Propaq CS monitor with the in-service
14 General information Welch Allyn Propaq CS Vital Signs MonitorNoteThe inservice mode is not available if the monitor detects that a sensor has been
215SetupPrepare the Propaq CS monitor for a new patient1. Press the gray recessed MONITOR button on the right side of the monitor to turn the monitor
ii Welch Allyn Propaq CS Vital Signs Monitor© 2008 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, t
16 Setup Welch Allyn Propaq CS Vital Signs MonitorWhenever you change the patient mode, the alarm limit settings, maximum NIBP cuff inflation pressure
Directions for Use Setup 17Set patient alarms and alarm limits1. From the Main Menu, press SETUP, ALARMS to access the Alarms Status Menu:2. Press LIM
18 Setup Welch Allyn Propaq CS Vital Signs Monitor3. Press NEXT PARAMETER to highlight the parameter you want to change, then press NEXT SETTING to hi
Directions for Use Setup 19Change Powerup Patient Mode or store customized settingsThe Propaq CS monitor has standard, preset, default powerup setting
20 Setup Welch Allyn Propaq CS Vital Signs MonitorCustomize Patient Mode settings1. From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the
321MonitoringPerform ECG/RESP monitoringWARNING Impedance pneumography detects respiratory effort via changes in chest volume; therefore, impedance
22 Monitoring Welch Allyn Propaq CS Vital Signs Monitor• Impedance pneumography (RESP) is not recommended for use with high frequency ventilation.• Si
Directions for Use Monitoring 23• Severe artifact and interference (such as defibrillation interference) can cause the waveform to move off the displa
24 Monitoring Welch Allyn Propaq CS Vital Signs Monitor7. Plug the ECG cable into the ECG connector on the monitor's left side panel.8. Support t
Directions for Use Monitoring 2513. Press MORE to display the second ECG/RESP menu and status window:014. Press NEXT and CHANGE as desired to adjust t
iiiContents1 - General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Intended use. . . . . . . . . . . . .
26 Monitoring Welch Allyn Propaq CS Vital Signs Monitor15. Set alarms according to your facility’s standards.Use the ECG filter to display a better wa
Directions for Use Monitoring 27Using the Propaq CS monitor with pacemaker patientsIf the patient being monitored has a pacemaker, the Propaq CS monit
28 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform Invasive Blood Pressure (IBP) monitoring1. Inspect the transducer cable and transducer
Directions for Use Monitoring 295. If the ZERO menu is not displayed, from the Main Menu press INVASIVE PRESSURE, and then ZERO P1 (or ZERO P2). The w
30 Monitoring Welch Allyn Propaq CS Vital Signs Monitor10. To display all invasive pressure waveforms on one scale (when two IBP channels are active),
Directions for Use Monitoring 3113. To change the format of the IBP numerics, from the second IBP menu press FORMAT.To restore the first format, press
32 Monitoring Welch Allyn Propaq CS Vital Signs MonitorTake a Noninvasive Blood-Pressure (NIBP) readingAt powerup, the Propaq CS monitor has an NIBP d
Directions for Use Monitoring 33To enable the Smartcuf filter:• Connect the ECG leads to the patient and perform ECG monitoring during NIBP.• From the
34 Monitoring Welch Allyn Propaq CS Vital Signs Monitor3. Place the cuff on the limb.If possible, place it at the same level as the heart. If above th
Directions for Use Monitoring 356. If motion artifact such as shivering, coughing, or other motion interferes with NIBP readings, do the following:Pos
iv Contents Welch Allyn Propaq CS Vital Signs Monitor7 - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
36 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform temperature monitoring1. Place the temperature probe on the patient.2. Plug the probe c
Directions for Use Monitoring 37Perform SpO2 monitoringEach SpO2 sensor is designed for application to a specific site on the patient within a certain
38 Monitoring Welch Allyn Propaq CS Vital Signs Monitorto the sensor, the patient is in cardiac arrest or shock, or the patient has hypotension, sever
Directions for Use Monitoring 39Perform SpO2 monitoring with Nellcor option1. Attach the sensor to the patient according to the sensor manufacturer&ap
40 Monitoring Welch Allyn Propaq CS Vital Signs Monitor9. If the C-LOCK function is desired, press C-LOCK to set it to ON.C-LOCK synchronizes the puls
Directions for Use Monitoring 41Perform SpO2 “spot-check” monitoringThe SpO2 Standby Mode allows you to remove the SpO2 sensor from a patient without
42 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform Mainstream CO2 monitoringThe capnography (CO2) option measures End-tidal CO2 (ETCO2), I
Directions for Use Monitoring 433. Plug in the CO2 sensor cable to the mainstream CO2 connector on the monitor left side panel.4. See page 46 and set
44 Monitoring Welch Allyn Propaq CS Vital Signs MonitorPerform sidestream CO2 monitoringThe capnography (CO2) option measures End-tidal CO2 (ETCO2), I
Directions for Use Monitoring 452. See page 46 and set up the CO2 display and alarm limits, and then continue this procedure with step 3.3. A. For a n
11General informationIntended useThe Propaq CS monitor is intended to be used by skilled clinicians for multiparameter vital signs monitoring of neona
46 Monitoring Welch Allyn Propaq CS Vital Signs MonitorSidestream CO2 accuracy decreases if additional tubing is connected to the sample line. Avoid c
Directions for Use Monitoring 472. Press RANGE to select the CO2 waveform scale or range.To ch a n g e C O2 units (mmHg, kPa, or %) see “Change the
48 Monitoring Welch Allyn Propaq CS Vital Signs Monitor7. P r e s s SOURCE to change between mainstream CO2 and sidestream CO2 monitoring (if both o
449Alarms & AlertsDescription of alarm and alert tone patterns• Patient alarm tone pattern: one second on, two seconds off• Apnea alarm tone patte
50 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorRe-enable an alarm or alert tone before the 90-second silence period has elapsed1. P r
Directions for Use Alarms & Alerts 51Inhibit alarm and alert tones for four minutes: 4 MIN SUSPNDAs you perform patient care, there might be occas
52 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorInhibit alarm and alert tones indefinitely: ALL ALARMSYou can suspend potential or cur
Directions for Use Alarms & Alerts 53Summary of alarm and alert keys and Acuity Central Station messagesThe following table summarizes Silence and
54 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorCustomize alarm limits based on patient’s current vital signsTo quickly set all alarm
Directions for Use Alarms & Alerts 55Alarm holdoffsTo help minimize false alarms, the monitor briefly delays or “holds off” triggering alarms for
2 General information Welch Allyn Propaq CS Vital Signs MonitorSymbolsThe following symbols may appear on the Propaq CS monitor or accessories. They a
56 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorTroubleshooting equipment alert messagesSimultaneous equipment alertsMULTIPLE. If mult
Directions for Use Alarms & Alerts 57IBP messagesTRANSDUCER NOT DETECTED. The transducer connection is broken.TRANSDUCER SHORT CIRCUIT. This messa
58 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitorchannel has not updated its calibration in 15 minutes, the channel will briefly deacti
Directions for Use Alarms & Alerts 59Temperature messagesPROBE NOT DETECTED. This message occurs when the Propaq CS monitor has successfully measu
60 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitorwaveform is displayed without range values. Replace the adapter or replace the sensor.
Directions for Use Alarms & Alerts 61ALTIMETER NOT CALIBRATED - EEPROM. The sidestream CO2 option has not been calibrated. Refer the Propaq CS mon
62 Alarms & Alerts Welch Allyn Propaq CS Vital Signs MonitorNetwork alert messages with AcuityCHECK ACUITY/SERIAL CONNECTION. The Propaq CS detect
Directions for Use Alarms & Alerts 63Printer alert messagesCHECK DOOR. The door on the bottom of the printer is open. Close the door to remove thi
64 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitor
565Printing & trendsPrint patient dataPrint the displayed waveformsPress SNAPSHOT or START/STOP. The Propaq CS monitor prints up to three of the d
Directions for Use General information 3General warnings and cautionsFamiliarize yourself with all warnings and cautions before using the Propaq CS mo
66 Printing & trends Welch Allyn Propaq CS Vital Signs MonitorDisplay or print trendsDisplay or print a single trend1. To display a patient data t
Directions for Use Printing & trends 67Print multiple trends manually or automatically1. P r e s s SETUP, MORE, PRINTER from the Main Menu to di
68 Printing & trends Welch Allyn Propaq CS Vital Signs MonitorSet printer options and automatic printing1. P r e s s SETUP, MORE, PRINTER from t
Directions for Use Printing & trends 69Print OxyCRGThe OxyCRG is a printout of two minutes of continuous HR/PR and SpO2 numerics, and a compressed
70 Printing & trends Welch Allyn Propaq CS Vital Signs Monitor
671AcuityUnderstanding the Propaq CS, Acuity, and FlexNet networkThis section describes the operation of the Propaq CS monitor (software version 3.5X)
72 Acuity Welch Allyn Propaq CS Vital Signs MonitorThe Propaq CS monitor can communicate with an Acuity Central Station as part of Welch Allyn’s FlexN
Directions for Use Acuity 73If the Wireless Propaq CS is moved out of range or loses communication with the FlexNet network and Acuity, it continues t
74 Acuity Welch Allyn Propaq CS Vital Signs MonitorThe Acuity MenuThe Acuity Menu on the Propaq CS allows you to control some of the interaction with
Directions for Use Acuity 75Hardwired (Serial) connection to AcuityConnect the monitor to the Acuity System1. If the Propaq CS monitor has already bee
4 General information Welch Allyn Propaq CS Vital Signs MonitorWARNING Do not connect more than one patient to a monitor. Do not connect more than o
76 Acuity Welch Allyn Propaq CS Vital Signs MonitorDisconnect the hardwired monitor from the Acuity System1. To disconnect the Propaq CS monitor from
Directions for Use Acuity 77Wireless connection to AcuityConnect a new patient1. Make sure the Acuity network cable is not plugged into the Acuity con
78 Acuity Welch Allyn Propaq CS Vital Signs MonitorMonitor a patient out of range of AcuityWhile out of range of Acuity, the Wireless Propaq CS contin
Directions for Use Acuity 79Reassign a monitored patient to a new room in the same unitIf a patient is being monitored and you want to assign them to
80 Acuity Welch Allyn Propaq CS Vital Signs MonitorReassign the Wireless Propaq CS to a new patientIf you want to discontinue monitoring a patient and
781MaintenanceAvoid electrostatic dischargeWhen humidity in the working environment decreases, the human body and other insulators can become charged
82 Maintenance Welch Allyn Propaq CS Vital Signs Monitor2. Plug the ac adapter power cord into the ac power adapter and the ac power source outlet.3.
Directions for Use Maintenance 83Replace power adapter fusesIf the green battery charging light is off and the ac power adapter does not provide power
84 Maintenance Welch Allyn Propaq CS Vital Signs MonitorReplace monitor input power fuseIf the green battery charging light is off and the ac power ad
Directions for Use Maintenance 85Install printer paper1. Lay the monitor on its back to gain access to the bottom of the printer.2. Squeeze the locks
Directions for Use General information 5WARNING If a product has been dropped or severely abused, send it to a qualified service person to confirm p
86 Maintenance Welch Allyn Propaq CS Vital Signs MonitorInspect and clean the monitor and accessoriesBefore cleaning, thoroughly inspect the monitor a
Directions for Use Maintenance 87Service interval recommendationsAt the intervals recommended below, qualified biomedical service personnel should ser
88 Maintenance Welch Allyn Propaq CS Vital Signs MonitorExtended storage precautionsRemoving the battery will erase all stored Custom patient mode set
Directions for Use Maintenance 89The monitor displays a confirmation screen asking you to confirm that you want to change the network name.• If you pr
90 Maintenance Welch Allyn Propaq CS Vital Signs Monitor
891ReferenceSet the time and date1. From the Main Menu press SETUP, MORE, MORE to display the Time/Day window:2. Press NEXT, UP, and DOWN as needed to
92 Reference Welch Allyn Propaq CS Vital Signs MonitorChange the date format, ECG filter, and units1. Make sure you are in the Adult patient mode (fro
Directions for Use Reference 93Factory default settingsThe monitor is shipped from the factory with these preset default settings. For information abo
94 Reference Welch Allyn Propaq CS Vital Signs MonitorHR Limits Adult: 50, 120 beats per minutePed: 50, 150 beats per minuteNeonate: 100, 200 beats pe
Directions for Use Reference 95SpecificationsECG specificationsThe ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and A