Welch Allyn MicroTymp 3 Handle Instruções de Operação Página 21

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 72
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 20
MicroTymp
2
Completing a Test
Figure 21
Tympanogram Section of Printout
The tympanogram is a graph which records the admittance of the ear as
a function of air pressure.
Data Section of Printout
The data section displays numeric values for the four key characteristics
of the tympanogram:
Peak Ya — the compensated static acoustic admittance (height) of
the peak, measured in acoustic millimhos (mmho).
Gradient (GR) — the width of the tympanogram; the distance across
the tympanogram measured at a height 50% down from the peak,
measured in decapascals (daPa).
Tympanic Peak Pressure (TPP) — where the tympanometric peak
occurred on the pressure axis, measured in decapascals (daPa).
Volume of the Ear Canal (Vea) — acoustically-determined ear canal
volume, measured in cubic centimeters (cc) at +200 daPa.
17
Tympanogram
Data
(Optional)
Interpretive
Message
(Optional)
Vista de página 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 71 72

Comentários a estes Manuais

Sem comentários