WelchAllynU.K.Ltd.
Cublington Road, Aston Abbotts, Buckinghamshire HP22 4ND, England
Tel: 0 207 365 6780 Fax: 0 207 365 9694
WelchAllynEuropeanCustomerService
Navan Business Park, Dublin Road, Navan, County Meath, Ireland
Tel: +353 46 906 7790 Fax: +353 46 906 7754
WelchAllynPTY.LTD,S.Africa
394 Surrey Ave, Ferndale, Randburg 2194 South Africa
P O Box 809, Fontainebleau, South Africa 2032
Tel: 0 11 777 7555 Fax: 0 11 777 7556
www.welchallyn.com
© 2009 Welch Allyn SM2517EU Rev D
All rights reserved. Acuity, Propaq and Smartcuf are registered trademarks and FlexNet is a trademark of Welch Allyn.
Other trademarks mentioned are for identification purposes only and are owned by their respective companies.
Welch Allyn Acuity
®
Central Monitoring
Station utilizes Aruba Mobile Edge
Architecture for best-in-class HIPAA security,
a superior voice and data network solution
and high availability for critical applications.
Only Masimo SET SpO
2
sensors, including
LNOP durable adhesive and NR value disposable
sensors, should be used with Masimo pulse oxi-
metry. Masimo, Masimo SET, LNOP, and NR are
registered trademarks of Masimo Corporation.
Only Nellcor
®
OxiMax
®
pulse oximetry sensors
should be used with the Nellcor pulse oximetry
option. NELLCOR and OxiMax are registered
trademarks of Nellcor Puritan Bennett, Inc.
All Propaq CS vital signs monitors come standard
with Smartcuf technology, a new standard in
NIBP accuracy. This patented software offers
industry-leading performance and accuracy in
the presence of motion artifacts.
Alarms
•
Allparameters: upper/lower limits
•
Allparameters:Adult/Paediatric/Neonate patient
mode-specific limits
•
Factory default or programmable settings for all
patient modes
•
Alarmindicator: red
•
Alarm(s)offindicator:yellow
•
Audiblealarmtone: high/med/low
•
Alarmsuspend:90 seconds
•
One-button Stat Set for all alarm limits
(except Apnea Delay)
•
EQAlert: yellow
Trends
•
Tabular numeric format
•
All parameters trended/viewable
•
Resolution:non-NIBP trends entered every 2 min
•
NIBP trends entered after each reading
•
Duration: 5 hours non-NIBP trends
(up to 150 readings)
8 hours NIBP trends
(up to 128 readings)
•
Page up/down trend view
Inservice Mode
•
Includes two sets of simulated patient data including
waveforms, for training and education
Colour Active Matrix Display
•
Type: TFT (Thin Film Transistor) LCD module
•
Resolution: 640 x 480 pixels, 1 pixel = R+G+B dots
•
Activeviewingarea: 170.9 x 129.6 mm
•
Pixelpitch: 0.267 mm
•
Viewingangle: U/D 40°, R/L 60° (typical), ≥ 10:1
contrast ratio
• Contrastratio:150:1 (typical); measured in dark
room at centre of screen
•
Displaycolour:18-bit (6 bits per primary color)
Waveformsandmatchingnumerics: Predefined
•
Luminance: 200 cd/m
2
(typical); measured at
saturation point
•
Responsetime: 40 ms (maximum); “white to black”
Environmental
•
Operatingtemperature:5 to 40 °C (monitor),
5 to 40 °C (printer)
•
Operatingrelativehumidity: 15% to 95%,
noncondensing (MIL STD 810E), printer 35%
to 85%, noncondensing (MIL STD 810E)
•
Operatingaltitude:-610 to 4,572 m
(817 to 429 mmHg)
•
Shipping/storagetemperature: -20 to 60 °C
•
Shock: 50 g (monitor); 30 g (expansion module/
printer)
•
Randomvibration:Designed to meet RTCA
DO-160D, category C (monitor)
•
Electromagneticcompatibility(EMC):
per EN60601-1-2, 1993
•
Drip-proof per IEC 529, level IPX1 (monitor)
Physical
•
Monitor(includinghandle):
Size: 20.8 cm (H) x 24.4 cm (W) x 14.1 cm (D)
Weight:3.4 kg
•
MonitorwithSpO
2
Module:
Size:20.8 cm (H) x 24.4 cm (W) x 19.7 cm (D)
Weight:4.9 kg
•
MonitorwithExpansionModule:
(with Printer/SpO
2
/MSCO
2
)
Size:28.8 cm (H) x 24.4 cm (W) x 19.7 cm (D)
Weight:6.5 kg
Power
•
Rechargeable sealed lead acid internal battery pack
•
Internal recharger
•
Inputvoltage:12 to 28 VDC, 25 W
•
ACpoweradapter: 100 to 120 VAC, 50 to 60 Hz
(North America/Japan); 220 to 240 VAC, 50 to
60 Hz (International)
•
Operatingtimeonbattery,typical*: Monitor
with SpO
2
option: 4 hours; Monitor with Expansion
Module (Printer, SpO
2
/CO
2
): 3 hours;
Monitor only:2 hours
•
Batteryrechargetime:6 to 8 hours
(when monitor OFF)
•
Batteryrechargetime:8 to 12 hours
(when monitor ON)
Certifications/Standards
•
Complies with relevant AAMI, IEC, EN, CSA
and UL standards. CE Marked according to
the European Medical Device Directive. CSA
Certificate of Compliance for use in the U.S.
and Canada
VA Contract V797P-3486k
DSCP Contract SP0200-97-D-8021
* Fully charged, new battery at 25 °C, NIBP and
Snapshot (if applicable) every 15 minutes, all
accessories used.
Comentários a estes Manuais