Insight Instruments Solarc™ Light SourceFor P/N: 20002.000 InsightInstruments InsightInstruments InsightInstruments InsightInstrumentsInsight Instrume
6 UI-20002 6/2002MountingSecure bracket to a flat surface with screws provided. Align mounting post on Light Source with mounting hole on bracket (Vie
Insight Instruments 7Lamp Replacement1. Turn unit off and unplug the power cord from both the wall outlet and the Insight Instruments Light Source.2.
8 UI-20002 6/20025. Reconnect the lamp to the lamp connector and place in the spring mechanism. Verify that the pin on the lamp is properly seated in
Insight Instruments 95. Reinsert the fuse holder by pressing it into the power connector until it snaps into place.6. If the new fuse blows, immediate
10 UI-20002 6/2002SpecificationsElectricalInput Voltage: 100 - 240 VAC, 50 - 60 HzInput Current: 1A (@ 100 VAC)Output: 60 VDC @ 0.4ALeakage Current: l
Insight Instruments 11Equipment ClassificationClass I, Type B, Continuous OperationFusesT1.0A L250V: 1.0A, 250V, time lag (slow-blow), low breaking ca
12 UI-20002 6/2002TroubleshootingIf lamp is not illuminating and...POSSIBLE CAUSESOLUTIONfan is not operating.Light Source is not plugged into either
Insight Instruments 13Maintenance And RepairFor minor trouble, refer to the troubleshooting section in this manual for possible causes and corrective
14 UI-20002 6/2002WarrantyInsight Instruments warrants the Insight Instruments Solarc™ Light Source when new, to be free of defects in material and wo
Insight Instruments, Inc.5400 S. Bryant Ave.Sanford, FL 32773, USATel.: 1-800-255-8354 or1-407-324-2727Fax: 1-407-324-2744Printed in the USA UI2002
Table of ContentsSymbol Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Indications . . . . . . . . . . . . . . . . . . . . . .
Insight Instruments 1Symbol DescriptionsAttention: Read Operating Manual for Warnings, Precautions, and Instructions for Use.Dangerous Voltage: Refer
2 UI-20002 6/2002WARNING: possible explosion hazard. Do not use in the presence of flammable anesthetics. To ensure patient and operator safety, read
Insight Instruments 3CAUTION: Only qualified medical personnel should use the Light Source.Provide adequate ventilation to prevent overheating. Do not
4 UI-20002 6/2002System DescriptionThe Insight Instruments Light Source is used to supply high quality, white light for video and optical illumination
Insight Instruments 5Set-up And OperationAfter a 15-second warm-up period, the light output will stabilize at its normal intensity. Plug in the fiber