18 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
Figure 6. 9600 Plus Calibration Tester with Braun ThermoScan Pro 3000
1. Clean the probe tip with a cotton swab slightly moistened
with isopropyl alcohol, remove excess alcohol with a clean
cotton swab, and let air dry for 5 minutes.Do not use any
chemical other than alcohol to clean the probe window.
2. Place the Braun ThermoScan Pro 3000 in calibration check
mode using the following steps.
a. Wait until the thermometer is in Standby mode (the screen
is blank).
b. Press down and hold the Activation button, do not release
until Step d.
c. Press the /Mem button.
Vista de página 23